Some examples below. These are NORMAL waveform segments. The the OA's and CA's would not count toward AHI in a lab sleep study.


***********************************************
Below is an example of a disturbance or arousal of unknown origin.
The machine scored a 16 second Obstructive Apnea (OA); however, we can see that there is sharp increase and irregularity in the Flow Rate Waveform immediately preceding the event, indicating a likely arousal. Additionally, we can see that the pulse rate and the oxygen level actually increase during the event. With an actual OA, it would be typical to see the O2 fall during the event and then rise with an increase in pulse after.
Likely, I changed positions or rolled over and held my breath for 16 seconds while doing so. The small break in the Pulse Oximeter trace reinforces the supposition that this was movement related.
This event would not be scored as an OA in a lab sleep study

***********************************************
CA Examples:

Edited to consolidate images