I asked this question before and I think it was Dori that gave me the answer I needed. I need to know how to change the pressure setting on my c-pap. I remember something about unplugging it first then plugging it back in but I forgot the rest. I have a Remstar pro cflex m series. Please help.
Laura
Change my pressure setting
- OceanGoingGal
- Posts: 326
- Joined: Tue Oct 07, 2008 3:51 am
- Location: Lakeland Florida
- Contact:
Change my pressure setting
"Fools Names And Fools Faces Are Often Seen In Public Places"............Adelaide B DiNardi
Ask Me About Girl Talk Cafe!
Ask Me About Girl Talk Cafe!
Re: Change my pressure setting
Hold the two "arrow" buttons down while you plug it in. It will boot in Clinician mode. From there, you can scroll thru the settings and change them.
The OSA patient died quietly in his sleep.
Unlike his passengers who died screaming as the car went over the cliff...
Unlike his passengers who died screaming as the car went over the cliff...
- OceanGoingGal
- Posts: 326
- Joined: Tue Oct 07, 2008 3:51 am
- Location: Lakeland Florida
- Contact:
Re: Change my pressure setting
Thank you. I appreciate the info.
Laura
Laura
"Fools Names And Fools Faces Are Often Seen In Public Places"............Adelaide B DiNardi
Ask Me About Girl Talk Cafe!
Ask Me About Girl Talk Cafe!
Re: Change my pressure setting
Nice to see you back Laura!
_________________
| Mask: AirFit™ P10 Nasal Pillow CPAP Mask with Headgear |
| Additional Comments: ResScan software 3.13, Pressure 21/15 |
“Life is 10% what happens to you, and 90% what you make of it.” Charles Swindoll
- rested gal
- Posts: 12880
- Joined: Thu Sep 09, 2004 10:14 pm
- Location: Tennessee
Re: Change my pressure setting
From a post here:
Accessing the setup menu - M series Auto machines
viewtopic.php?p=305612#p305612
To go into the Setup menu of the Respironics REMstar Auto|Aflex M series machine:
1. Unplug the machine at the back.
2. Lift the little lid on top of the machine.
3. Press both the right/left arrow buttons and keep holding both arrow buttons down together while you plug the machine in again. Continue holding those two buttons down after you plug it in. You'll hear two beeps. Keep holding those right/left arrow buttons down. You'll hear a single beep. (Or just count off ten seconds from the time you plugged it in while holding the two arrow buttons.) The setup menu is unlocked now.
4. You'll see the word "Setup" in the lower right corner of the little window on the machine.
Press the " + " button under the word "Setup." That opens the therapy setup menu.
At this point, I'd STRONGLY suggest that you do NOT make any changes to the settings yet. Get a pencil and paper. You might want to get a magnifying glass, too!! Go through the entire menu, writing down every word, every symbol, and every number you see in the window for each menu item. That way you'll have a record of the original settings if you change something later and then decide you want to go back to the original settings.
5. To go through the menu, use the right/left arrow buttons as if they were forward/backward keys.
6. To make a change in a setting, use the plus and minus buttons as if they were up/down keys.
Any changes you make are automatically saved when you exit from the menu.
Very Important: If you come to a setting that says "Split Night Time" (the options for it are "off 120 180 240") ALWAYS keep that setting turned OFF.
7. To exit the menu, briefly press the middle button (with the circle symbol on it) that would normally start the motor blowing.
_____________________________
Accessing the setup menu - M series Auto machines
viewtopic.php?p=305612#p305612
To go into the Setup menu of the Respironics REMstar Auto|Aflex M series machine:
1. Unplug the machine at the back.
2. Lift the little lid on top of the machine.
3. Press both the right/left arrow buttons and keep holding both arrow buttons down together while you plug the machine in again. Continue holding those two buttons down after you plug it in. You'll hear two beeps. Keep holding those right/left arrow buttons down. You'll hear a single beep. (Or just count off ten seconds from the time you plugged it in while holding the two arrow buttons.) The setup menu is unlocked now.
4. You'll see the word "Setup" in the lower right corner of the little window on the machine.
Press the " + " button under the word "Setup." That opens the therapy setup menu.
At this point, I'd STRONGLY suggest that you do NOT make any changes to the settings yet. Get a pencil and paper. You might want to get a magnifying glass, too!! Go through the entire menu, writing down every word, every symbol, and every number you see in the window for each menu item. That way you'll have a record of the original settings if you change something later and then decide you want to go back to the original settings.
5. To go through the menu, use the right/left arrow buttons as if they were forward/backward keys.
6. To make a change in a setting, use the plus and minus buttons as if they were up/down keys.
Any changes you make are automatically saved when you exit from the menu.
Very Important: If you come to a setting that says "Split Night Time" (the options for it are "off 120 180 240") ALWAYS keep that setting turned OFF.
7. To exit the menu, briefly press the middle button (with the circle symbol on it) that would normally start the motor blowing.
_____________________________
ResMed S9 VPAP Auto (ASV)
Humidifier: Integrated + Climate Control hose
Mask: Aeiomed Headrest (deconstructed, with homemade straps
3M painters tape over mouth
ALL LINKS by rested gal:
viewtopic.php?t=17435
Humidifier: Integrated + Climate Control hose
Mask: Aeiomed Headrest (deconstructed, with homemade straps
3M painters tape over mouth
ALL LINKS by rested gal:
viewtopic.php?t=17435