E-15 error code

General Discussion on any topic relating to CPAP and/or Sleep Apnea.
mouse

E-15 error code

Post by mouse » Sat Dec 13, 2008 5:13 am

HELP HAVE A REMstar c-FLEX REMPlus cpap showing an error code of E-15 can anyone tell me what this is and help me get a good nights sleep.Thank You

User avatar
6PtStar
Posts: 2659
Joined: Mon Jun 18, 2007 1:58 pm
Location: Texas, Is there any other place?

Re: E-15 error code

Post by 6PtStar » Sat Dec 13, 2008 9:22 am

The list I have shows E-15 as:

E-15 ERR_ARM_UNDEFINED_INSTRUCTION_EXCEPTION •Defective FLASH chip •Replace Main PCA

I am not sure what a flash chip but the fix of replacing the main circuit board does not sound good. Sorry!

Jerry

_________________
Humidifier: HC150 Heated Humidifier With Hose, 2 Chambers and Stand
Additional Comments: 11cm/H2O, Encore Pro 1.8i, Pro Analyzer, Encore Viewer1.0 - 3 Remstar Pro2's, 1 Remstar Auto
Life's journey is not to arrive at the grave safely in a well preserved body, but rather to skid in sideways, totally worn out, shouting: "Wow what a ride!"
I still play Cowboys and Bad Guys but now I use real bullets. CAS

Pinhead
Posts: 85
Joined: Sat Apr 19, 2008 5:14 am
Location: The Villages Fl

Re: E-15 error code

Post by Pinhead » Sat Dec 13, 2008 10:26 am

I think that is the code I had on mine. It was under warrenty so they fixed it free. It appears that they changed much of the insides. The DME said that often this problem is caused by moving it with water in it. If your is out of warrenty or your insurance won't replace you might be better off to get a new one on line then get it fixed I would think it would be cheaper.

Take it into your DME when you can and they can tell you in about five minutes if they have to send it in or not.

Jack

User avatar
Wulfman
Posts: 12321
Joined: Thu Jul 07, 2005 3:43 pm
Location: Nearest fishing spot

Re: E-15 error code

Post by Wulfman » Sat Dec 13, 2008 10:45 am

Actually, the E-15 error code on the REMstar Plus w/C-Flex shows this in the Service Manual:

Error code: E-15
Problem: Rotor locked. The motor could not be restarted.
Corrective Action: Check wiring. Replace blower assembly. If error persists, replace main PCA